Validation of the AmpFlSTR® Identifiler® Direct PCR Amplification kit in a laboratory accredited according to the ISO17025 standard

Research output: Contribution to journalJournal articleResearchpeer-review

A total of 249 FTA® card samples were investigated with the Identifiler® Direct kit. PCR amplification was performed according to the manufacturer's protocol with modifications in the reaction volume (10-25μL), the number of PCR cycles (25-27) and the number of FTA-card punches (1-2). Full concordance was observed between the results obtained with the Identifiler® Direct kit and the washed punches typed with the Identifiler® kit. The success rate with one 1.2mm punch in a 15μL PCR reaction volume and 25 PCR-cycles was 93.3%. After retyping of the failed samples, the combined success rate was 98.7%. Overall, the quality of the results was reduced when the number of punches was increased or the PCR reaction volume was decreased. The Identifiler® Direct master mix may be used in combination with other primer sets for direct amplification of other loci, e.g. the 49 SNPs using the SNPforID multiplex assay.

Original languageEnglish
JournalForensic Science International: Genetics Supplement Series
Volume3
Issue number1
Pages (from-to)e165-e166
ISSN1875-1768
DOIs
Publication statusPublished - Dec 2011

    Research areas

  • Crime DNA reference samples, FTA card, Identifiler Direct Amplification kit, SNPforID multiplex assay, STR typing

ID: 307818588