Shadow effects in simulated ultrasound images derived from computed tomography images using a focused beam tracing model

Research output: Contribution to journalJournal articleResearchpeer-review

Simulation of ultrasound images based on computed tomography (CT) data has previously been performed with different approaches. Shadow effects are normally pronounced in ultrasound images, so they should be included in the simulation. In this study, a method to capture the shadow effects has been developed, which makes the simulated ultrasound images appear more realistic. The method using a focused beam tracing model gives diffuse shadows that are similar to the ones observed in measurements on real objects. Ultrasound images of a cod (Gadus morhua) were obtained with a BK Medical 2202 ProFocus ultrasound scanner (BK Medical, Herlev, Denmark) equipped with a dedicated research interface giving access to beamformed radio frequency data. CT images were obtained with an Aquilion ONE Toshiba CT scanner (Toshiba Medical Systems Corp., Tochigi, Japan). CT data were mapped from Hounsfield units to backscatter strength, attenuation coefficients, and characteristic acoustic impedance. The focused beam tracing model was used to create maps of the transmission coefficient and scattering strength maps. Field II was then used to simulate an ultrasound image of 38.9 × 55.3 × 4.5 mm, using 10 6 point scatterers. As there is no quantitative method to assess quality of a simulated ultrasound image compared to a measured one, visual inspection was used for evaluation.

Original languageEnglish
JournalJournal of the Acoustical Society of America
Volume132
Issue number1
Pages (from-to)487-497
Number of pages11
ISSN0001-4966
DOIs
Publication statusPublished - Jul 2012

ID: 331500207