Pressure Difference Estimation in Non-stenotic Carotid Bifurcation Phantoms Using Vector Flow Imaging

Research output: Contribution to journalJournal articleResearchpeer-review

Documents

  • Fulltext

    Accepted author manuscript, 223 KB, PDF document

Local pressure differences estimated using vector flow imaging (VFI) and direct catheterization in seven carotid bifurcation phantoms were compared with simulated pressure fields. VFI correlated strongly with simulated peak pressure differences (r = 0.99, p < 0.00001), with an average overestimation of 12.3 Pa (28.6%). The range between the lowest and highest pressure difference of VFI underestimated simulations by 4.6 Pa (8.06%; r = 0.99, p < 0.0001). The catheter method exhibited no correlation (r = –0.09, p = 0.85). Ten repeated measurements on one phantom revealed a small standard deviation (SD) for VFI (SD = 8.4%, mean estimated SD = 11.5%), but not for the catheter method (SD = 785.6%). An in vivo peak systolic pressure difference of 97.9 Pa (estimated SD = 30%) was measured using VFI in one healthy individual. This study indicates that VFI pressure difference estimation is feasible in phantoms and in vivo and realistic estimates of the SD can be attained from the data.

Original languageEnglish
JournalUltrasound in Medicine and Biology
Volume48
Issue number2
Pages (from-to)346-357
ISSN0301-5629
DOIs
Publication statusPublished - 2022

Bibliographical note

Publisher Copyright:
© 2021 World Federation for Ultrasound in Medicine & Biology

    Research areas

  • Carotid bifurcation, Computational fluid dynamics, Fluid-filled pressure catheter, Fluid–structure interaction, Pressure difference estimation, Ultrasound, Vector flow imaging