Bone density measurements adjacent to acetabular cups in total hip arthroplasty using dual-energy CT: an in vivo reliability and agreement study
Research output: Contribution to journal › Journal article › Research › peer-review
Documents
- Open Access Version
Final published version, 900 KB, PDF document
Background: Measuring bone mineral density (BMD) around acetabular prosthetic components with computed tomography (CT) is challenged by the complex anatomy and metal artifacts. Three-dimensional (3D) segmentation is required for the analysis, but it is usually not practically applicable on current CT workstations.
Purpose: To test the between-scan agreement and reliability of custom segmentation software for BMD measurements adjacent to cemented and uncemented acetabular cups in dual-energy CT (DECT).
Material and Methods: Twenty-four male patients with total hip arthroplasty were scanned and rescanned using 130-keV virtual monochromatic DECT images. Hemispherical regions of interest were defined slice-by-slice and BMD was calculated around the acetabular cup using custom segmentation software.
Results: In the uncemented cup, the mean BMD was 153 mg/cm3 with a between-scan difference of 10 mg/cm3 (P < 0.0001). In the cemented cup, the mean BMD was 186 mg/cm3 with a between-scan difference of 6 mg/cm3 (P = 0.15). In both uncemented and cemented cups the intraclass correlation coefficient between repeated measurements was >0.95 and narrow Bland-Altman Limits of Agreement.
Conclusion: BMD can be measured with high absolute between-scan agreement and good reliability adjacent to acetabular cemented and uncemented cups using DECT and segmentation software.