A novel lumen-apposing metal stent for endoscopic ultrasound-guided drainage of pancreatic fluid collections: a prospective cohort study

Research output: Contribution to journalLetterResearchpeer-review

  • Daisy Walter
  • Uwe Will
  • Andres Sanchez-Yague
  • Dirk Brenke
  • Jochen Hampe
  • Helge Wollny
  • Jose Miguel Esteban López-Jamar
  • Gertrud Jechart
  • Vilmann, Peter
  • Joan B Gornals
  • Sebastian Ullrich
  • Martin Fähndrich
  • Alberto Herreros de Tejada
  • Félix Junquera
  • Ferran Gonzalez-Huix
  • Peter D Siersema
  • Frank P Vleggaar

BACKGROUND AND STUDY AIMS: A novel large-diameter, lumen-apposing, self-expanding metal stent with bilateral flanges was recently developed for endoscopic ultrasound (EUS)-guided transmural drainage of symptomatic pancreatic fluid collections (PFCs). The aim of this study was to evaluate the efficacy and safety of this stent in a large cohort.

PATIENTS AND METHODS: Patients with a PFC undergoing EUS-guided drainage with this novel stent were prospectively enrolled in this multicenter cohort study.

RESULTS: There were 61 patients: 46 patients (75 %) with walled-off necrosis (WON) and 15 (25 %) with a pancreatic pseudocyst. Stent placement was technically successful in 60 patients (98 %, 95 %CI 95 % - 100 %). Clinical success, defined as resolution of clinical symptoms in combination with a decrease in the PFC size to ≤ 2 cm on imaging, was achieved in 93 % of patients with a pancreatic pseudocyst (95 %CI 77 % - 100 %) and in 81 % of patients with WON (95 %CI 69 % - 94 %). Treatment failure occurred in nine patients (16 %, 95 %CI 6 % - 26 %), including four patients who required surgical intervention. Stent removal was performed in 82 % of patients after a median of 32 days (range 2 - 178) and was rated as easy in all but one patient. In 10 patients, endoscopic stent removal was not performed because of stent migration (n = 3), stent dislodgement during necrosectomy (n = 3), stent removal during surgery (n = 2), or refusal by the patient (n = 2). In total, five major complications were reported (9 %, 95 %CI 2 % - 16 %), including PFC infection (n = 4) and perforation (n = 1).

CONCLUSION: EUS-guided drainage using this novel stent is feasible and the clinical results obtained are promising with a low major complication rate.

Original languageEnglish
JournalEndoscopy
Volume47
Issue number1
Pages (from-to)63-67
Number of pages5
ISSN0013-726X
DOIs
Publication statusPublished - Jan 2015

    Research areas

  • Adult, Aged, Drainage, Endosonography, Female, Follow-Up Studies, Humans, Logistic Models, Male, Middle Aged, Pancreatic Pseudocyst, Pancreatitis, Acute Necrotizing, Prospective Studies, Stents, Treatment Outcome, Ultrasonography, Interventional

ID: 152932319