Validation of a high-performance liquid chromatography/fluorescence detection method for the simultaneous quantification of fifteen polycyclic aromatic hydrocarbons

Research output: Contribution to journalJournal articleResearchpeer-review

A high-performance liquid chromatography/fluorescence method using multiple wavelength shift for simultaneous quantification of different PAH compounds was developed. The new method was superior to the methods of DONG and GREENBERG [J. Liquid Chromatogr. 11, 1887-1905 (1988)] and WISE et al. [Polycyclic aromat. Hydrocarb. (in press)] with respect to sensitivity of detection of the majority of 15 PAH compounds, and in particular of naphthalene, acenaphthene, fluorene and benzo(b)fluoranthene. The method of validation analysis employed showed that the new method is in statistical balance meaning that no systematic errors, and only small unsystematic errors, could be demonstrated. Furthermore, the method had a good reproducibility and a high sensitivity.
Original languageEnglish
JournalAnnals of Occupational Hygiene
Volume35
Issue number6
Pages (from-to)603-11
Number of pages9
ISSN2398-7308
Publication statusPublished - Dec 1991

    Research areas

  • Air Pollutants, Occupational, Chromatography, High Pressure Liquid, Evaluation Studies as Topic, Fluorescent Dyes, Humans, Polycyclic Compounds, Reproducibility of Results

ID: 48866083