Intracranial Pressure: A Comparison of the Noninvasive HeadSense Monitor versus Lumbar Pressure Measurement

Research output: Contribution to journalJournal articleResearchpeer-review

OBJECTIVE: To compare a new method of noninvasive intracranial pressure (nICP) measurement with conventional lumbar puncture (LP) opening pressure.

METHODS: In a prospective multicenter study, patients undergoing LP for diagnostic purposes underwent intracranial pressure measurements with HeadSense, a noninvasive transcranial acoustic device, and indirectly with LP. Noninvasive measurements were conducted with the head in a 30° tilt and in supine position before and after LP. The primary endpoint was the correlation between nICP measurement in supine position before LP and the LP opening pressure.

RESULTS: There was no correlation between supine nICPs before LP and the LP opening pressures (r = -0.211, P = 0.358). The 30° head-tilt nICPs correlated with the supine nICPs before LP (r = 0.830, P < 0.01). There was no correlation between supine nICPs before and after LP (r = 0.056, P = 0.831) or between 30° head-tilt nICPs and LP opening pressures (r = -0.038, P = 0.861).

CONCLUSIONS: There was no correlation between nICPs and LP opening pressures. Further development is warranted before transcranial acoustic HeadSense can become a clinical tool for investigating patients with neurologic conditions.

Original languageEnglish
JournalWorld Neurosurgery
Volume112
Pages (from-to)e576-e580
ISSN1878-8750
DOIs
Publication statusPublished - 2018

    Research areas

  • Adult, Aged, Female, Humans, Intracranial Hypertension/diagnosis, Intracranial Pressure/physiology, Male, Middle Aged, Monitoring, Physiologic/methods, Patient Positioning, Prospective Studies, Spinal Puncture, Young Adult

ID: 215866503