Assessment of a portable device for the quantitative measurement of ankle joint stiffness in spastic individuals

Research output: Contribution to journalJournal articleResearchpeer-review

OBJECTIVE: Spasticity is a common complication with neurological diseases and CNS lesions. Instrumented systems to evaluate spasticity often cannot provide an immediate result, thus limiting their clinical usefulness. In this study we investigated the accuracy and reliability of the portable Neurokinetics RA1 Ridgidity Analyzer to measure stiffness of the ankle joint in 46 controls, 14 spinal cord injured (SCI) and 23 multiple sclerosis (MS) participants. METHODS: Ankle stiffness measures were made twice by two raters, at speeds above and below the expected stretch reflex threshold. Ankle torque was measured with the portable device and a stationary torque motor. Inter- and intra-rater reliability was assessed with the intra-class correlation coefficient (ICC). RESULTS: Stiffness measures with the portable and stationary devices were significantly correlated for controls and MS participants (p
Original languageEnglish
JournalClinical Neurophysiology
Volume123
Issue number7
Pages (from-to)1371-1382
Number of pages12
ISSN1388-2457
DOIs
Publication statusPublished - 2012

    Research areas

  • Spasticity, Spasticity evaluation, Stiffness, Spinal cord injury, Multiple sclerosis

ID: 36076296